11 Safe-Drive™
157
4 1 Turn the Retractor 60 degrees clockwise (i.e. to
the right) to lock the bayonet.
2 Push down the latch on the Outer casing to
secure the connecon.
3 Insert safety pin.
5 1 Close the blow-out ball valve underneath the
Isolaon valve.
2 Li up the Isolaon valve handle locking plate.
3 Open the Isolaon valve by turning the valve
handle 90 degrees counterclockwise. The valve is
open when the ball valve handle is parallel to the
Retractor and sensor.
6 Now the sensor can be inserted into the process.
Turn the hand-wheel counterclockwise unl it
stops, i.e. unl the sensor flange connects with the
Isolaon valve and only the end of the screw thread
is visible.
Warning! If you detect leaking, revert immediately
to the previous step. Do not connue installaon
unl the reason for leakage has been cleared and
fixed.
7 Fit the four M12 nuts to the bolts holding the
sensor to the Isolaon valve and screw them on
with a 19 mm or 3/4" wrench.
Important: Do not ghten the nuts too hard, set
the torque at 50 Nm (37 /lbs).
8 1 Turn the wheel 90 degrees (a quarter turn) to the
right.
2 Remove the safety pin.
3 Unlock the latch of the Outer casing.
4 Turn the casing 60 degrees counterclockwise i.e.
unl the handle is up on top.